time:2020-12-11 browse:second
PS10SR-2388 pressure transducer is produced by inserting a piezoresistive sensor into stainless steel housing with differential pressure ports. The measured pressure acts on silicon sensitive element through pressure port, transmitting pressure into voltage output signal by linearity relationship. Thick film circuit by laser trimming compensates zero and temperature performance of sensor, the product is widely applied in Pressure calibration device, Automobile, Aviation, Spaceflight, Air pressure control and Industrial measure, etc. Especially for transducer with pressure port HB4-3, it has been widely used for aviation industry.
·Compensate zero and FS thermal error by laser trimming;
·Used for non-conductive and non-corrosive liquid or gas;
·Optional constructions;
·Compact size, easy installation;
·Gauge, Absolute and Sealed gauge;
·Constant current supply;
Pressure Range | 0~0.02MPa…7MPa |
Power Supply | 1.5mA DC |
Over-Pressure | 1.5 times FS |
Zero output | ≤ ±2 mV DC |
FS Output | ≥ 50mV DC |
Accuracy | ±0.25%FS(typical) ±0.50%FS(max.) |
Zero Thermal Drift | ±1.0%FS/℃(typical) ±2.0%FS/℃(max.) |
FS Thermal Drift | ±1.0%FS/℃(typical) ±2.0%FS/℃(max.) |
Compensation temp. | 0℃~50℃ |
Operation temp. | -40℃~120℃ |
Long-term stability | ±0.2%FS/year(typical) ±0.3%FS/year(max.) |
(Unit: mm)
Order Guide | |||||||
PS10SR-2388 | Piezoresistive Pressure Transducer | ||||||
Code | Pressure range | Remark | Code | Pressure range | Remark | ||
0B | 0kPa~20kPa | G | 09 | 0kPa~700kPa | G、A | ||
0A | 0kPa~35kPa | G | 10 | 0MPa~1.0MPa | G、A | ||
02 | 0kPa~70kPa | G、A | 12 | 0MPa~2.0MPa | G、A | ||
03 | 0kPa~100kPa | G、A | 13 | 0MPa~3.5MPa | G、A、S | ||
07 | 0kPa~200kPa | G、A | 14 | 0MPa~7.0MPa | S | ||
08 | 0kPa~350kPa | G、A | |||||
Code | Pressure type | ||||||
G | Gauge | ||||||
A | Absolute | ||||||
S | Sealed gauge | ||||||
Code | Power Supply | ||||||
L | Constant current supply | ||||||
Code | Pressure port (customized ports are also optional) | ||||||
I | M12×1 male(HB4-3), Code I (with 1.5m cable) | ||||||
II | M12×1 male(flat head), Code II (plug connection) | ||||||
III | M10 male(ball head), Code III (Φ0.45 Kovar) | ||||||
PS10SR-2388 | 08 | G | L | I | the whole spec. |
Order Note
1. 1.5m cable is free of charge; the extra cable should be paid additionally;
2. For your special use or requirements, please contact with our company freely.
PS10SR-2388 piezoresistive pressure transducer is widely applied in Pressure calibration device, Automobile, Aviation, Spaceflight, Air pressure control and Industrial measure, etc. Especially for transducer with pressure port HB4-3, it has been widely used for aviation industry.